Measurement Modes Details.

Mode 1: This is the most common mode of measurement. The sample's temperature is adjusted to the set target temperature and the measurement is taken.

Mode 2: This mode is used to measure the sample quickly and when accuracy is not as critical. After the START key is pressed, the unit measures the Refractive Index and temperature at short intervals and diplays the estimated value at the target temperature.

Mode 3: This mode is the quickest way to measure and is best suited when the properties of the sample are changing constantly. In this mode, the unit starts measuring immediately after the START key is pressed. A time delay can also be programmed to postpone the measurement after pressing START.

Mode S: (RX-5000i only) This mode is for emulsion samples. It displays the measurement value once a certain level of sample stability is achieved..





 